510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0609-2016Class II

Siemens Acute Care CTNI TestPak-in vitro diagnostic test for the measurement of cardiac troponin I in...

Siemens Healthcare Diagnostics Inc. / initiated 2015-11-11 / Terminated / root cause: labeling and UDI

Siemens Healthcare Diagnostics Inc began a recall of Siemens Acute Care CTNI TestPak-in vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma Catalog Number: CCTNI SMN: 10445071 on . FDA classified it as Class II and gives the reason as "Stratus CS Cartridge Barcode labels may not scan on the Stratus CS systems". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0609-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Acute Care CTNI TestPak-in vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma Catalog Number: CCTNI SMN: 10445071
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
Stratus CS Cartridge Barcode labels may not scan on the Stratus CS systems
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on November 11, 2015 both in the United States and internationally for communication with affected customers. This notice informs customers of the issue and provides mitigation instructions. Customers with questions were advised to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding call 781-269-3000.
Quantity in commerce
5558
Distribution
Worldwide Distribution - US (nationwide), Algeria, Armenia, Australia, Austria, Bangladesh, Bosnia Herzegovina, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Libya, Malaysia, Poland, Portugal, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., and United Kingdom.
Product codes and lots
Lot Numbers: 235151002, 235159002, 235166002, 235173002, 235208002 235215002, 235222002, 235229002, 235236002, 235243002 235250002, 235257002, 235278002, 235285002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0609-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0609-2016Class IITerminatedVoluntary: Firm initiated
Z-0610-2016Class IITerminatedVoluntary: Firm initiated
Z-0611-2016Class IITerminatedVoluntary: Firm initiated
Z-0612-2016Class IITerminatedVoluntary: Firm initiated
Z-0613-2016Class IITerminatedVoluntary: Firm initiated
Z-0614-2016Class IITerminatedVoluntary: Firm initiated
Z-0615-2016Class IITerminatedVoluntary: Firm initiated
Z-0616-2016Class IITerminatedVoluntary: Firm initiated
Z-0617-2016Class IITerminatedVoluntary: Firm initiated
Z-0618-2016Class IITerminatedVoluntary: Firm initiated
Z-0619-2016Class IITerminatedVoluntary: Firm initiated