Z-0764-2016Class II

Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications

Philips North America LLC / initiated 2015-11-03 / Terminated / root cause: design

Philips Electronics North America Corporation began a recall of Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications on . FDA classified it as Class II and gives the reason as "Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0764-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
Root cause tag
design
FDA root cause
Device Design
Action
The letters for the customers in the United States were sent on 11/3/2015, via certified return receipt and were addressed to the Risk Manager. Philips will perform a test on the 15 affected Allura Xper R8.2 FD 20C systems to see if the motion controller amplifier of the brake motor works correctly. If the test will not pass, the motor will be replaced. The Recall Package is sent to all Key Markets worldwide. The Key Market of an affected country will contact the facility where these systems were installed.
Quantity in commerce
15 subject to correction and removal
Distribution
Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.
Product codes and lots
Ceiling version only.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0764-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0764-2016Class IITerminatedVoluntary: Firm initiated
Z-0765-2016Class IITerminatedVoluntary: Firm initiated