Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications
Philips Electronics North America Corporation began a recall of Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications on . FDA classified it as Class II and gives the reason as "Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0764-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Allura Xper FD20 Ceiling version only; 722028; imaging applications.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The letters for the customers in the United States were sent on 11/3/2015, via certified return receipt and were addressed to the Risk Manager. Philips will perform a test on the 15 affected Allura Xper R8.2 FD 20C systems to see if the motion controller amplifier of the brake motor works correctly. If the test will not pass, the motor will be replaced. The Recall Package is sent to all Key Markets worldwide. The Key Market of an affected country will contact the facility where these systems were installed.
- Quantity in commerce
- 15 subject to correction and removal
- Distribution
- Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.
- Product codes and lots
- Ceiling version only.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0764-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0765-2016 | Class II | Terminated | Voluntary: Firm initiated |