Z-0359-2016Class II

Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal...

ConMed Corporation / initiated 2015-11-05 / Terminated / root cause: manufacturing process

ConMed Corporation began a recall of Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source on . FDA classified it as Class II and gives the reason as "Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal"...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0359-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On November 4, 2015, ConMed Corporation distributed Urgent Device Recall Notification letters dated November 5, 2015 to their customers via overnight courier service. Customers should stop the use of these devices immediately. The recall notice recipient should contact all organizations within their facility and any other facilities that may have received the affected products. It is imperative that all end users of these devices receive this notice and respond immediately. Customers with questions can call Patricia Cotter, ConMed Recall Coordinator 315-624-3237 (8AM - 5PM EST, Monday through Friday), or fax to 315-624-3225 or email [email protected]. Customers who have any of the devices subject to recall (listed on the recall notice-attachment I) should immediately contact their sales representative. The sales representative will deliver a replacement unit to your facility and will pick up and return the recalled devices to ConMed Corporation. If you are unable to reach your sales representative, please contact Patricia Cotter at the numbers above. If you do not have any devices to return, customers should complete the Business Reply Form (attachment II), indicating you have no devivces and fax it to 315-624-3225, Attn: Patricia Cotter.
Quantity in commerce
9 devices
Distribution
Distributed in the states of CA, and TX and in the country of Mexico.
Product codes and lots
Serial Numbers: 1) 15090103 & 15150005, 15150006, 15150007, 15150008, 15150009, 15150010, 15150011, 15150012, 15130004, 15250013, 15250014, 15250015, 15250016, 15250017, 15300018 & 15300020, 2) 15050001, 15300005 & 15300006

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0359-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0359-2016Class IITerminatedVoluntary: Firm initiated