510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0641-2016Class II

VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750009916...

Ortho-Clinical Diagnostics, Inc. / initiated 2015-11-03 / Terminated / root cause: software

Ortho-Clinical Diagnostics began a recall of VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750009916, Software Product Code 6802864, Software Version 3.2 and... on . FDA classified it as Class II and gives the reason as "Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0641-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750009916, Software Product Code 6802864, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.
Root cause tag
software
FDA root cause
Software design
Action
Ortho-Clinical Diagnostics sent an Urgent Product Correction Letters (Ref. CL2015-211, dated November 3, 2015) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address) to all customers who received the affected devices (VITROS 3600 and 5600 Systems). The notification identified the affected products, problem and actions to be taken. Foreign affiliates were informed via e-mail on 11/03/2015 of the issue and instructed to notify their consignees of the issue and required actions. Ortho-Clinical Diagnostics sent an Urgent Product Correction Letters (Ref. CL2015-221, dated November 11, 2015) via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address) to all customers who received the affected devices (VITROS 5,1 FS and 4600 users). The notification identified the affected products, problem and actions to be taken. Foreign affiliates were informed via e-mail on 11/11/2015 of the issue and instructed to notify their consignees of the issue and required actions.
Quantity in commerce
Domestic: 1083, Foreign: 999
Distribution
Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.
Product codes and lots
Software Version 3.2 and below; Serial Numbers J56000024 - J56002309

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0641-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0640-2016Class IITerminatedVoluntary: Firm initiated
Z-0641-2016Class IITerminatedVoluntary: Firm initiated
Z-0642-2016Class IITerminatedVoluntary: Firm initiated
Z-0643-2016Class IITerminatedVoluntary: Firm initiated