Z-0377-2016Class II
AMS 4675 Angio Pack. For use in a general clinical procedure
Windstone Medical Packaging, Inc. / initiated 2015-11-02 / Terminated / root cause: manufacturing process
Windstone Medical Packaging, Inc. began a recall of AMS 4675 Angio Pack. For use in a general clinical procedure on . FDA classified it as Class II and gives the reason as "AMS 4675 Angio Pack is recalled because this kit containing U-SDN 2 Part Ultra Modified Seldinger Needle which is recalled by supplier". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0377-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Windstone Medical Packaging, Inc.
- Product
- AMS 4675 Angio Pack. For use in a general clinical procedure.
- Product code
- KDD Kit, Surgical Instrument, Disposable
- Reason for recall
- AMS 4675 Angio Pack is recalled because this kit containing U-SDN 2 Part Ultra Modified Seldinger Needle which is recalled by supplier.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Aligned Medical Solutions (AMS) sent the Procedure Products -Urgent Medical Device Recall Letter, dated November 02, 2015 to consignees. This letter listed the items affected by recall as: AMS 4674 Custom Pack and AMS 4675 Angio Pack. Both direct account and sub account are informed that the part # NS/37 -0387 01 U-SDN 2 Part Ultra Modified Seldinger Needle (introducer needle) is recalled because some was unable to accept guide wires up to .038" as intended. The inner diameter was found to be out of specification in some cases. All other components in these two AMS kits are not affected by this recall. Customers with questions about this recall can call Vicki Davis at 321-527-7714 or send email to [email protected].
- Quantity in commerce
- 40 kits
- Distribution
- Distributed in AL and VA.
- Product codes and lots
- Lot 84371, expiration date: 2/28/2017
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0376-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0377-2016 | Class II | Terminated | Voluntary: Firm initiated |