510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0426-2016Class II

Myla Master Dvd 3.2 Cli Dl380

bioMerieux, Inc. / initiated 2015-11-03 / Terminated / root cause: software

bioMerieux, Inc. began a recall of MYLA MASTER DVD 3.2 CLI DL380 Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments... on . FDA classified it as Class II and gives the reason as "MYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0426-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
MYLA MASTER DVD 3.2 CLI DL380 Product Usage: MYLA is a computer application (Middleware) based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s)).
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
MYLA software connected to a Laboratory Information System (LIS) and a VITEK 2 system has the potential to link a test result to an incorrect patient record with the same specimen ID.
Root cause tag
software
FDA root cause
Software design
Action
bioMerieux sent an Urgent Product Correction Notice letter dated November10, 2015 via FedEx to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local bioMrieux Clinical Customer Service organization at (800) 682-2666.
Quantity in commerce
626 systems in total
Distribution
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and countries of: Foreign accounts: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, France, Germany, Greece, Hungary, India, Italy, Japan, Korea, Netherlands, Poland, Portugal, Russia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, Angola, Puerto Rico, Republic of Belarus, Republic of Belarus, Benin, Bosnia, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroon, Chad, Democratic Republic of Congo, Costa Rica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Gabon, Guam, Guatemala, Guatemala, Guinea, Honduras, Indonesia, Iraq, Israel, Ivory Coast, Kazakhstan, Kenya, Kenya, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Malaysia, Malaysia, Mali, Portugal, Myanmar, Managua, Niger, Nigeria, Palestine, Philippines, Philippines, Qatar, Kingdom of Saudi Arabia, Senegal, Serbia, Slovakia, Togo, Tunisia, Ukraine, UAE, Vietnam, Vietnam, Ivory Coast and Zimbabwe.
Product codes and lots
REF # 415358, Serial number/Software version V3.2 CLI

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0426-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0425-2016Class IITerminatedVoluntary: Firm initiated
Z-0426-2016Class IITerminatedVoluntary: Firm initiated
Z-0427-2016Class IITerminatedVoluntary: Firm initiated
Z-0428-2016Class IITerminatedVoluntary: Firm initiated
Z-0429-2016Class IITerminatedVoluntary: Firm initiated
Z-0430-2016Class IITerminatedVoluntary: Firm initiated
Z-0431-2016Class IITerminatedVoluntary: Firm initiated