Z-0389-2016Class II

LeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 6F, 80 cm, Model 1601-68. Indicated for the...

LeMaitre Vascular Inc. / initiated 2015-10-21 / Terminated / root cause: packaging

LeMaitre Vascular, Inc. began a recall of LeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 6F, 80 cm, Model 1601-68. Indicated for the removal of arterial emboli and thrombi on . FDA classified it as Class II and gives the reason as "Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0389-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LeMaitre Vascular Inc.
Product
LeMaitre NovaSil Silicone Single Lumen Embolectomy Catheter 6F, 80 cm, Model 1601-68. Indicated for the removal of arterial emboli and thrombi.
Product code
DXE Catheter, Embolectomy
Reason for recall
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Consignees were notified of the recall on October 21, 2015 via letter. Consignees were asked to do the following: Identify all catheters with the affected LOT and model numbers in inventory; quarantine all unused catheters from the affected LOTs and record this action in the attached form; send the form via mail, email or fax to Customer Service who will then issue an RMA-number (Return Merchandise Authorization number) for the return shipment and replacement of the products; pass this notice on to all those who need to be aware within your organization and to any organization to which the potentially affected devices have been transferred. For questions concerning this safety notice, please contact Salva Kozin, Quality Assurance Engineer, at 781-221-2266 ext. 183.
Quantity in commerce
4315 total
Distribution
Worldwide Distribution. US nationwide including PUERTO RICO, and Australia, AUSTRIA, BELGIUM, COLOMBIA, COSTA RICA, Czech Republic, DENMARK, EGYPT, FRANCE, GERMANY, ICELAND, Italy, JAPAN, JORDAN, Latvia, LUXEMBOURG, MOROCCO, NETHERLANDS, Norway, Oman, PERU, PORTUGAL, Poland, Slovenia, Spain, Switzerland, Sweden, THAILAND, United Kingdom, United Arab Emirates, and VIETNAM.
Product codes and lots
Lot Number: NSE1663

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0389-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0378-2016Class IITerminatedVoluntary: Firm initiated
Z-0379-2016Class IITerminatedVoluntary: Firm initiated
Z-0380-2016Class IITerminatedVoluntary: Firm initiated
Z-0381-2016Class IITerminatedVoluntary: Firm initiated
Z-0382-2016Class IITerminatedVoluntary: Firm initiated
Z-0383-2016Class IITerminatedVoluntary: Firm initiated
Z-0384-2016Class IITerminatedVoluntary: Firm initiated
Z-0385-2016Class IITerminatedVoluntary: Firm initiated
Z-0386-2016Class IITerminatedVoluntary: Firm initiated
Z-0387-2016Class IITerminatedVoluntary: Firm initiated
Z-0388-2016Class IITerminatedVoluntary: Firm initiated
Z-0389-2016Class IITerminatedVoluntary: Firm initiated
Z-0390-2016Class IITerminatedVoluntary: Firm initiated