510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0372-2016Class II

Smith & Nephew PERI-LOC VOLAR DISTAL RADIUS PLATE STANDARD 3 HOLE SHORT, LEFT - 62 MM, STERILE R, Rx only...

Smith & Nephew, Inc. / initiated 2015-10-20 / Terminated / root cause: manufacturing process

Smith & Nephew, Inc. began a recall of Smith & Nephew PERI-LOC VOLAR DISTAL RADIUS PLATE STANDARD 3 HOLE SHORT, LEFT - 62 MM, STERILE R, Rx only, REF Numbers: 71863110 Bone fixation plate on . FDA classified it as Class II and gives the reason as "The devices were packaged and laser marked anatomically incorrect. The Left plates were laser marked and packaged as Right plates and the Right plates were laser marked and packaged as Left plates". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0372-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew PERI-LOC VOLAR DISTAL RADIUS PLATE STANDARD 3 HOLE SHORT, LEFT - 62 MM, STERILE R, Rx only, REF Numbers: 71863110 Bone fixation plate
Product code
HRS Plate, Fixation, Bone
Reason for recall
The devices were packaged and laser marked anatomically incorrect. The Left plates were laser marked and packaged as Right plates and the Right plates were laser marked and packaged as Left plates.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Distributors, notified by Fed- Ex on 10/20/2015, were asked to quarantine the product and return it.
Quantity in commerce
9 units
Distribution
International only Germany, Japan, Russia
Product codes and lots
Batch # 14KT47574

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0372-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0372-2016Class IITerminatedVoluntary: Firm initiated
Z-0373-2016Class IITerminatedVoluntary: Firm initiated