Z-0564-2016Class II

Radiographic Transparency for Certain and External Connection Tapered Implants. Images that assist with...

ZimVie US Corp LLC / initiated 2015-11-16 / Terminated / root cause: software

Biomet 3i, LLC began a recall of Radiographic Transparency for Certain and External Connection Tapered Implants. Images that assist with measurements for dental implants on . FDA classified it as Class II and gives the reason as "Images contained within radiographic transparency ART 1036 are undersized". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0564-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Radiographic Transparency for Certain and External Connection Tapered Implants. Images that assist with measurements for dental implants.
Product code
NDP Accessories, Implant, Dental, Endosseous
Reason for recall
Images contained within radiographic transparency ART 1036 are undersized.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
BIOMET3i sent an" Urgent Medical Device Recall letter " dated November, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Responsibilities: 1. Please review this notice and check your inventory for the affected item by comparing any on-hand ART 1036 Transparencies to a standard 8.5" x 11" sheet of paper. The non-conforming ART 1036 is approximately 7.8" x 10.2". 2. Immediately quarantine and remove all affected items from service. 3. Complete the attached Business Reply Form and either: a. fax it to 561-514-6316 or b. email it to [email protected] or c. return it along with the affected product using the included shipping label 4. Return affected product to BIOMET 3i. For assistance or other questions that you may have relative to this notice, please contact BIOMET 3i at 1- 800-342-5454 or 1-561-776-6700. Team members are available to assist you 8:00am to 6:00pm (Eastern), Monday through Friday.
Quantity in commerce
686
Distribution
Worldwide Distribution - US Distribution including Puerto Rico., and to the states of : CA, CO, FL, IL, MA, MI, MO, NY, OH, SC, TX, VA, WI, WV and Hawaii., and to the countries of : Australia, Germany, France, Greece, Northern Mariana Islands, Taiwan and Uruguay.
Product codes and lots
Model Number: ART 1036, Rev. C 03/13; Purchase Order # 184427.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0564-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0564-2016Class IITerminatedVoluntary: Firm initiated