Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0540-2016Class III

RealityCHECK Tilt Cup 12 Panel + Adult, Item No. RC12TCA

Ameditech, Inc. / initiated 2015-11-16 / Terminated / root cause: manufacturing process

Ameditech Inc began a recall of RealityCHECK Tilt Cup 12 Panel + Adult, Item No. RC12TCA on . FDA classified it as Class III and gives the reason as "Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0540-2016
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ameditech, Inc.
Product
RealityCHECK Tilt Cup 12 Panel + Adult, Item No. RC12TCA
Product code
JXM Enzyme Immunoassay, Benzodiazepine
Reason for recall
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
A Medical Device Labeling Correction letter was sent on 11/16/15 to distributors and end users to inform them that Ameditechs Drugs of Abuse Tests have shown reduced reactivity for up to two (2) of the nineteen (19) BZO compounds for which specificity information is provided in the product insert. The letter informs the customers that the particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. Customers with questions are instructed to contact Ameditech at (844) 616-4453 or email at [email protected].
Quantity in commerce
80 kits
Distribution
Nationwide Distribution.
Product codes and lots
153709

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0540-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0484-2016Class IIITerminatedVoluntary: Firm initiated
Z-0485-2016Class IIITerminatedVoluntary: Firm initiated
Z-0486-2016Class IIITerminatedVoluntary: Firm initiated
Z-0487-2016Class IIITerminatedVoluntary: Firm initiated
Z-0488-2016Class IIITerminatedVoluntary: Firm initiated
Z-0489-2016Class IIITerminatedVoluntary: Firm initiated
Z-0490-2016Class IIITerminatedVoluntary: Firm initiated
Z-0491-2016Class IIITerminatedVoluntary: Firm initiated
Z-0492-2016Class IIITerminatedVoluntary: Firm initiated
Z-0493-2016Class IIITerminatedVoluntary: Firm initiated
Z-0494-2016Class IIITerminatedVoluntary: Firm initiated
Z-0495-2016Class IIITerminatedVoluntary: Firm initiated
Z-0496-2016Class IIITerminatedVoluntary: Firm initiated
Z-0497-2016Class IIITerminatedVoluntary: Firm initiated
Z-0498-2016Class IIITerminatedVoluntary: Firm initiated
Z-0499-2016Class IIITerminatedVoluntary: Firm initiated
Z-0500-2016Class IIITerminatedVoluntary: Firm initiated
Z-0501-2016Class IIITerminatedVoluntary: Firm initiated
Z-0502-2016Class IIITerminatedVoluntary: Firm initiated
Z-0503-2016Class IIITerminatedVoluntary: Firm initiated
Z-0504-2016Class IIITerminatedVoluntary: Firm initiated
Z-0505-2016Class IIITerminatedVoluntary: Firm initiated
Z-0506-2016Class IIITerminatedVoluntary: Firm initiated
Z-0507-2016Class IIITerminatedVoluntary: Firm initiated
Z-0508-2016Class IIITerminatedVoluntary: Firm initiated
Z-0509-2016Class IIITerminatedVoluntary: Firm initiated
Z-0510-2016Class IIITerminatedVoluntary: Firm initiated
Z-0511-2016Class IIITerminatedVoluntary: Firm initiated
Z-0512-2016Class IIITerminatedVoluntary: Firm initiated
Z-0513-2016Class IIITerminatedVoluntary: Firm initiated
Z-0514-2016Class IIITerminatedVoluntary: Firm initiated
Z-0515-2016Class IIITerminatedVoluntary: Firm initiated
Z-0516-2016Class IIITerminatedVoluntary: Firm initiated
Z-0517-2016Class IIITerminatedVoluntary: Firm initiated
Z-0518-2016Class IIITerminatedVoluntary: Firm initiated
Z-0519-2016Class IIITerminatedVoluntary: Firm initiated
Z-0520-2016Class IIITerminatedVoluntary: Firm initiated
Z-0521-2016Class IIITerminatedVoluntary: Firm initiated
Z-0522-2016Class IIITerminatedVoluntary: Firm initiated
Z-0523-2016Class IIITerminatedVoluntary: Firm initiated
Z-0524-2016Class IIITerminatedVoluntary: Firm initiated
Z-0525-2016Class IIITerminatedVoluntary: Firm initiated
Z-0526-2016Class IIITerminatedVoluntary: Firm initiated
Z-0527-2016Class IIITerminatedVoluntary: Firm initiated
Z-0528-2016Class IIITerminatedVoluntary: Firm initiated
Z-0529-2016Class IIITerminatedVoluntary: Firm initiated
Z-0530-2016Class IIITerminatedVoluntary: Firm initiated
Z-0531-2016Class IIITerminatedVoluntary: Firm initiated
Z-0532-2016Class IIITerminatedVoluntary: Firm initiated
Z-0533-2016Class IIITerminatedVoluntary: Firm initiated
Z-0534-2016Class IIITerminatedVoluntary: Firm initiated
Z-0535-2016Class IIITerminatedVoluntary: Firm initiated
Z-0536-2016Class IIITerminatedVoluntary: Firm initiated
Z-0537-2016Class IIITerminatedVoluntary: Firm initiated
Z-0538-2016Class IIITerminatedVoluntary: Firm initiated
Z-0539-2016Class IIITerminatedVoluntary: Firm initiated
Z-0540-2016Class IIITerminatedVoluntary: Firm initiated
Z-0541-2016Class IIITerminatedVoluntary: Firm initiated
Z-0542-2016Class IIITerminatedVoluntary: Firm initiated