Z-0415-2016Class II

Catalog No. INF0020-A EnteraLite Infinity Enteral Pump Delivery Set with ENFit Connector System and...

Zevex, Inc. / initiated 2015-11-20 / Terminated

MOOG Medical Devices Group began a recall of Catalog No. INF0020-A EnteraLite Infinity Enteral Pump Delivery Set with ENFit Connector System and Transitional Stepped Connector (spiked set); Catalog No... on . FDA classified it as Class II and gives the reason as "Potential for leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0415-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zevex, Inc.
Product
Catalog No. INF0020-A EnteraLite Infinity Enteral Pump Delivery Set with ENFit Connector System and Transitional Stepped Connector (spiked set); Catalog No. INF0500-A - 500ml EnteraLite Infinity Enteral Pump Delivery Set with ENFit Connector System and Transitional Stepped Connector; Catalog No INF1200-A - 1200ml EnteraLite Infinity Enteral Pump Delivery Set with ENFit Connector System and Transitional Stepped Connector; Catalog No GR1200-A 1200ml Enteral Feeding Delivery Set with ENFit Connector System and Transitional Stepped Connector (gravity set). Used to dispense liquid nutrients at a preprogrammed pump or user controlled rate.
Product code
PIF Gastrointestinal Tubes With Enteral Specific Connectors
Reason for recall
Potential for leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector.
FDA root cause
Unknown/Undetermined by firm
Action
MOOG Medical sent an URGENT Field Safety Notice, dated November 20, 2015, to its customer. The letter identified the affected device, provided a description of the problem, and discussed the risk to health. The customer was asked to contact all known distributors and inform them that MOOG Medical will cease production of all its enteral administration sets using ENFit connectors and transition back to the previous revision of the product codes that do not include the ENFit connector (INF0020, INF0500, INF1200, and GR1200). MOOG Medical will produce the previous revision until a solution to the problem is found. All necessary users should be made aware of the notice. Christopher Dodge, Manager, Regulatory Affairs, should be contacted with questions or for additional info at 801-264-1001, ext. 112 or [email protected].
Quantity in commerce
12,014,130 sets
Distribution
Distributed to the state of MN.
Product codes and lots
All EnteraLite Infinity Enteral Pump Delivery Sets with ENFit connector System and transitional Stepped Connector.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0415-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0415-2016Class IITerminatedVoluntary: Firm initiated