FMS neXtra 3.5mm Aggressive Cutter Shaver Blades
DePuy Mitek, Inc., a Johnson & Johnson Co. began a recall of FMS neXtra 3.5mm Aggressive Cutter Shaver Blades; Product Code: 283315 Intended to provide controlled cutting, shaving, and abrading of tissue during... on . FDA classified it as Class II and gives the reason as "Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0657-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Mitek, Inc.
- Product
- FMS neXtra 3.5mm Aggressive Cutter Shaver Blades; Product Code: 283315 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.
- Product code
- HRX Arthroscope
- Reason for recall
- Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On November 20, 2015, customers were notified of the recall via letter. They were asked to take the following actions: Customer immediate actions: 1. Review inventory located within your facility and quarantine any products listed in this product recall notification. Please follow the instructions on the Business Reply Form to return any affected inventory located within your facility; 2. Review, complete, sign and return the attached business reply form to Elizabeth Messana ([email protected]) in accordance with the directions on the business reply form within 5 business days of receipt of this notification; 3. Forward this notice to anyone in your facility that needs to be informed; 4. If any potentially affected product has been forwarded to another facility, contact Elizabeth Messana at 508-828-3150; 5. Keep a copy of this notice.
- Quantity in commerce
- 45 units
- Distribution
- Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.
- Product codes and lots
- Lot Number: M1505053
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0656-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0657-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0658-2016 | Class II | Terminated | Voluntary: Firm initiated |