Cyberonics Vagus Nerve Stimulator VNS Therapy AspireSR Generator, Model 106
Cyberonics, Inc began a recall of Cyberonics Vagus Nerve Stimulator VNS Therapy AspireSR Generator, Model 106 on . FDA classified it as Class II and gives the reason as "Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0645-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Cyberonics Vagus Nerve Stimulator VNS Therapy AspireSR Generator, Model 106
- Product code
- LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
- Reason for recall
- Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- Cyberonics sent out written recall notification beginning October 14, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For further questions, please call 866-882-8804
- Quantity in commerce
- 29 units
- Distribution
- US Distributions to the states of : IL, NC, KS, and TX., and Internationally to the country of: United Kingdom.
- Product codes and lots
- Serial numbers 36602, 37019, 39097, 40311, 40465
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0645-2016 | Class II | Terminated | Voluntary: Firm initiated |