510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0566-2016Class II

Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The...

Aesculap Inc / initiated 2015-11-17 / Terminated / root cause: sterility

Aesculap, Inc. began a recall of Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the... on . FDA classified it as Class II and gives the reason as "AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0566-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.
Product code
GAD Retractor
Reason for recall
AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier
Root cause tag
sterility
FDA root cause
Mixed-up of materials/components
Action
The firm, Aesculap, sent an "URGENT MEDICAL DEVICE CORRECTION OR REMOVAL" letter dated 11/17/2015 to its customer. The letter described the product, problem and actions to be taken. The customer was instructed to review the product recall notification in its entirety; do not destroy any affected product; if you have no inventory remaining, notate that on the Product Removal Acknowledgement Form; contact our customer service department for an RGR at 866-229-3002 or email: [email protected] and immediately return affected product; complete and return the Product Removal Acknowledgement Form to Aesculap Quality Assurance department by fax to 610-791-6882 or email to [email protected] within two(2) weeks of receipt, even if the total inventory in your possession is zero (0) Sales Representative will make an appointment to come and exchange the effected product with an interium solution (FC360-Atrium Lift Retractor Set) that will allow any scheduled or unexpected cases, which would require the use of this product. The newly manufactured product will ship automatically when it's available.Then a visit will be scheduled to retrieve the loaner FC360R sets. If you have any questions, contact Quality Specialist at 610-984-9414 or email to [email protected] or Product Manager at 610-984-9150 or email to [email protected].
Quantity in commerce
41
Distribution
US Distribution to state of: NY.
Product codes and lots
Product Information All lots affected  Product Description FC424R Valve XS Atrium Lift Ret. Blade50X35 FC425R Valve XS Atrium Lift Ret. Blade60X35 FC426R Valve XS Atrium Lift Ret. Blade70X35 FC428R Valve XS Atrium Lift Ret. Blade50X45 FC429R Valve XS Atrium Lift Ret. Blade60X45 FC430R Valve XS Atrium Lift Ret. Blade70X45 FC431R Valve XS Atrium Lift Ret. Blade25X20 FC432R Valve XS Atrium Lift Ret. Blade25X25 FC433R Valve XS Atrium Lift Ret. Blade40X25 FC434R Valve XS Atrium Lift Ret. Blade50X25 FC435R Valve XS Atrium Lift Ret. Blade25X40 FC436R Valve XS Atrium Lift Ret. Blade25X50 FC420R Valve XS Atrium Lift Ret. Blade Complete Set

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0566-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0566-2016Class IITerminatedVoluntary: Firm initiated