510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0447-2016Class II

MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery

Elekta Limited / initiated 2015-11-24 / Terminated / root cause: software

Elekta, Inc. began a recall of MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery on . FDA classified it as Class II and gives the reason as "First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0447-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.
Root cause tag
software
FDA root cause
Software design
Action
Elekta sent an Urgent Important Field Safety Notice 371-01-MSQ-004 to all affected customers on 11/24/2015. The notice informs users to the issue, the work around and the solution. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta Office or Representative, as soon as possible but no longer than 30 days from receipt.
Quantity in commerce
211
Distribution
Nationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, & Puerto Rico.
Product codes and lots
Versions prior to 2.60.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0447-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0447-2016Class IITerminatedVoluntary: Firm initiated