Z-0461-2016Class II

Reagent Management System (RMS) Dimension RxL Max HM STM

Siemens Healthcare Diagnostics Inc. / initiated 2015-11-10 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics Inc. began a recall of Reagent Management System (RMS) Dimension RxL Max HM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical... on . FDA classified it as Class II and gives the reason as "Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0461-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Reagent Management System (RMS) Dimension RxL Max HM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Medical Device Correction (UMDC # DI-16-01.A.US) and Urgent Field Safety Notice (UFSN # DI-16-01.A.OUS) letters dated November 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. For questions contact your local Siemens technical support representative.
Quantity in commerce
151
Distribution
US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.
Product codes and lots
Dimension RXL MAX HM STM Catalog Numbers 752000.941 (Domestic)/752000.951 (Intl). RMS for RxL Catalog Numbers 753000.901 (Domestic)/753000.902 (Intl). All serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0461-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0458-2016Class IITerminatedVoluntary: Firm initiated
Z-0459-2016Class IITerminatedVoluntary: Firm initiated
Z-0460-2016Class IITerminatedVoluntary: Firm initiated
Z-0461-2016Class IITerminatedVoluntary: Firm initiated