Z-0588-2016Class II
Boston Scientific ACUITY Pro Coronary Sinus Guide Catheter, 9 Fr. Model 8104, 8105, 8106, 8107, 8108, 8109...
Boston Scientific Corporation / initiated 2015-11-25 / Terminated / root cause: manufacturing process
Boston Scientific Corporation began a recall of Boston Scientific ACUITY Pro Coronary Sinus Guide Catheter, 9 Fr. Model 8104, 8105, 8106, 8107, 8108, 8109, 8110, 8111, 8112, 8113, 8114, 8115, 8116, 8117... on . FDA classified it as Class II and gives the reason as "Potential for separation of the proximal handle from the shaft". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0588-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific ACUITY Pro Coronary Sinus Guide Catheter, 9 Fr. Model 8104, 8105, 8106, 8107, 8108, 8109, 8110, 8111, 8112, 8113, 8114, 8115, 8116, 8117, 8118, and 8119. The ACUITY Pro Lead Delivery System is designed for venous use to aid in the selective placement of cardiac resynchronization therapy (CRT) implantable venous leads in the cardiac vasculature. The catheter shafts are comprised of a PTFE inner liner, a reinforcing layer of stainless steel braid, and an outer polymer jacket. The distal end has a radiopaque polymer tip, while the proximal end has a hub with flush luer fitting to allow flush, contrast injection and aspiration polymer .
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Potential for separation of the proximal handle from the shaft.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Boston Scientific personnel began retrieving product from hospital shelves on November 25, 2015. The hospital was given a Medical Device Retrieval letter. The letter provided the reason for product retrieval, identified affected product, asked for product to be segregated to ensure that it will not be used, and to contact Boston Scientific for a return authorization number to ship product back to Boston Scientific. A Certificate of Removal was asked to be completed and emailed to [email protected] Questions can be addressed to Technical Services at 1-800-227-3422.
- Quantity in commerce
- 900 units
- Distribution
- Worldwide Distribution - US (nationwide) and Internationally to Argentina, Australia, Austria, Belgium, Colombia, Cyprus, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Portugal, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, and Thailand.
- Product codes and lots
- Model 8104, Lots: 18279548, 18279734, 18324877, 18329863, 18330946, 18348755, 18351569, 18353214, 18426640, 18461104, 18461521. Model 8105, Lots: 18281110, 18315176, 18322343, 18323156, 18323651, 18324468, 18327210, 18327359, 18345738, 18346540, 18346668, 18347522, 18352159, 18354329, 18379847, 18380498, 18393785, 18394034, 18395315, 18408497, 18408902, 18410018, 18428314, 18435809, 18437015, 18437680, 18439474, 18440388, 18444811, 18445630, 18446236, 18462103, 18462459, 18463154, 18466449, 18478544. Model 8106, Lots: 18286550, 18410783. Model 8107, Lots: 18285401, 18285605, 18312411, 18312981, 18313754, 18337534, 18338037, 18354248, 18411116, 18413022, 18413155, 18413834. Model 8108, Lots: 18273724, 18284294, 18338159, 18414856. Model 8109, Lots: 18285220, 18293137, 18294062, 18313856, 18314250, 18332118, 18337848, 18353117, 18415202, 18415923, 18416945, 18418278, 18419899. Model 8110, Lots: 18421013, 18473202. Model 8111, Lots: 18285930, 18351033, 18353540, 18420652, 18468909 . Model 8112, Lots: 18291473, 18336768, 18355225, 18421690, 18427953. Model 8113, Lots: 18324876, 18324963, 18332278, 18352696, 18358192, 18371973, 18372577, 18440783, 18441697, 18467134, 18468171, 18471598, 18472178, 18478892, 18479477. Model 8114, Lots: 18282143, 18284604, 18335173, 18345117, 18441212, 18442550, 18442887. Model 8115, Lots: 18288704, 18328790, 18329233, 18356691, 18357371, 18357372, 18372837, 18373201, 18377504, 18378867, 18446665, 18447045, 18451382, 18451950, 18452785, 18453099, 18464410, 18473932, 18477470, 18483223, 18483940. Model 8116, Lots: 18281865, 18473055. Model 8117, Lots: 18333920, 18334332, 18466601. Model 8118, Lots: 18316161, 18336837, 18348081. Model 8119, Lots: 18286265, 18291252, 18332911, 18336838, 18347898, 18480844, 18483020, 18471580.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0588-2016 | Class II | Terminated | Voluntary: Firm initiated |