Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray...
Siemens Medical Solutions USA, Inc began a recall of Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures on . FDA classified it as Class II and gives the reason as "Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0481-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis systems with large display from Siemens Medical Solutions USA, Inc. Interventional Fluoroscopic X-ray for angiographic procedures.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in factory screening.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent an Important Customer Safety Notice dated October 29, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The issue will be remedied with a hardware update AX025/15/S. The affected bypass module hardware will be exchanged with unaffected production lots to eliminate the problem. For further questions call: (610) 219-6300.
- Quantity in commerce
- 17
- Distribution
- US Distribution to the states of : IL, WI, MA, FL, AL, TN, ND, OH, TX, CA, NJ and MI.
- Product codes and lots
- Model numbers 10094135, 10094137, 10094141, 10848280, 10848281, 10848282, 10848355. Serial Numbers 123026, 147909, 123031, 103030, 147912, 147910, 147921, 147919, 131005, 121096, 125004, 137551, 121088, 109104, 109105, 109107, and 121085.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0481-2016 | Class II | Terminated | Voluntary: Firm initiated |