Covidien Endo GIA Ultra Universal 12 mm Short Stapler Handle Ref: EGIAUSHORT
Medtronic began a recall of Covidien Endo GIA Ultra Universal 12 mm Short Stapler Handle Ref: EGIAUSHORT Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and... on . FDA classified it as Class II and gives the reason as "Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0819-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Endo GIA Ultra Universal 12 mm Short Stapler Handle Ref: EGIAUSHORT Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
- Product code
- GDW Staple, Implantable
- Reason for recall
- Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medtronic sent an Urgent Medical Device Recall letters dated December 2 2015 and an Updated Urgent Medical Device Recall letter dated January 13, 2016, to affected customers, via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473. For questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.
- Quantity in commerce
- 364,768 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Albania, ALGERIA, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Greece, Guatemala, Honduras, Hungary, Iceland, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mauritius, Mexico, Monaco, Montenegro, Morocco, Namibia, Netherlands, Nicaragua, Nigeria, Norway, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, UAE, Ukraine, United Kingdom, Uruguay, Venezuela, and Zimbabwe.
- Product codes and lots
- RANGES OF LOT NUMBERS (all lot numbers within each range are affected) Lot # starting with P3C through P5B0289X P5C0196X through P5C0197X P5C0304X through P5C0307X P5C0796X through P5C0798X P5D0049X through P5D0051X P5D0146X through P5D0148X P5E0261X through P5E0262X P5F0158X through P5F0161X INDIVIDUAL LOT NUMBERS (in addition to ranges) N3F0260RX N3F0266RX N3F0267RX P5C0201X P5D0538X P5E0680X P5E0682X P5E0686X P5E0760X P5E0762X P5E0765X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0819-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0820-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0821-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0822-2016 | Class II | Terminated | Voluntary: Firm initiated |