Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL
Medtronic began a recall of Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA... on . FDA classified it as Class II and gives the reason as "Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0821-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
- Product code
- GDW Staple, Implantable
- Reason for recall
- Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medtronic sent an Urgent Medical Device Recall letters dated December 2 2015 and an Updated Urgent Medical Device Recall letter dated January 13, 2016, to affected customers, via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473. For questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.
- Quantity in commerce
- 327,797 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of Canada, Albania, ALGERIA, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Greece, Guatemala, Honduras, Hungary, Iceland, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mauritius, Mexico, Monaco, Montenegro, Morocco, Namibia, Netherlands, Nicaragua, Nigeria, Norway, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, UAE, Ukraine, United Kingdom, Uruguay, Venezuela, and Zimbabwe.
- Product codes and lots
- Lot numbers within each range are affected: Lot # starting with P3C through P5B0454X, P5B0468X, P5C0152X through P5E0450X. Individual Lot Number: P5B0468X, P5E0523X, P5E0273X
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0819-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0820-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0821-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0822-2016 | Class II | Terminated | Voluntary: Firm initiated |