510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0821-2016Class II

Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL

Medtronic, Inc. / initiated 2015-12-02 / Terminated / root cause: manufacturing process

Medtronic began a recall of Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA... on . FDA classified it as Class II and gives the reason as "Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0821-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Endo GIA Ultra Universal 12 mm XL Stapler Handle Ref: EGIAUXL Product Usage: The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. The Endo GIA Ultra Universal Short, Endo GIA Ultra Universal and Endo GIA Ultra Universal XL staplers when used with the Endo GIA curved tip single use reloads can be used to blunt dissect or separate target tissue from other tissue.
Product code
GDW Staple, Implantable
Reason for recall
Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Medtronic sent an Urgent Medical Device Recall letters dated December 2 2015 and an Updated Urgent Medical Device Recall letter dated January 13, 2016, to affected customers, via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473. For questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at (800) 882-5878.
Quantity in commerce
327,797 units
Distribution
Worldwide Distribution - US Nationwide and the countries of Canada, Albania, ALGERIA, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Georgia, Germany, Greece, Greece, Guatemala, Honduras, Hungary, Iceland, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mauritius, Mexico, Monaco, Montenegro, Morocco, Namibia, Netherlands, Nicaragua, Nigeria, Norway, Panama, Paraguay, Peru, Poland, Portugal, Romania, Russia, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, UAE, Ukraine, United Kingdom, Uruguay, Venezuela, and Zimbabwe.
Product codes and lots
Lot numbers within each range are affected: Lot # starting with P3C through P5B0454X, P5B0468X, P5C0152X through P5E0450X.  Individual Lot Number: P5B0468X, P5E0523X, P5E0273X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0821-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0819-2016Class IITerminatedVoluntary: Firm initiated
Z-0820-2016Class IITerminatedVoluntary: Firm initiated
Z-0821-2016Class IITerminatedVoluntary: Firm initiated
Z-0822-2016Class IITerminatedVoluntary: Firm initiated