Z-0590-2016Class II

NexGEN System REF 5980-37-01 Tibial Component Size Stemmed 3 Precoat, and REF 5986-37-01 Nonaugmentable...

Zimmer Manufacturing B.V. / initiated 2015-12-03 / Terminated / root cause: labeling and UDI

Zimmer Manufacturing B.V. began a recall of NexGEN System REF Tibial Component Size Stemmed 3 Precoat, and REF Nonaugmentable Tibial Component Option CR/PS/LPS Size Stemmed 3... on . FDA classified it as Class II and gives the reason as "A product complaint was received stating that a carton labeled as a Precoat Stemmed Tibia Size 3 was opened for use and the contents were a non-augmented Stemmed Tibia Plate size 3". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0590-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer Manufacturing B.V.
Product
NexGEN System REF 5980-37-01 Tibial Component Size Stemmed 3 Precoat, and REF 5986-37-01 Nonaugmentable Tibial Component Option CR/PS/LPS Size Stemmed 3. Tibial component of implanted knee prosthesis.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
A product complaint was received stating that a carton labeled as a Precoat Stemmed Tibia Size 3 was opened for use and the contents were a non-augmented Stemmed Tibia Plate size 3.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Zimmer Biomet sent an "Urgent Medical Device Recall-Lot Specific" letter to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to do the following: Review the notification and ensure affected personnel are aware of the contents. 2. Locate all affected product identified above and quarantine them immediately. 3. Carry out a physical count of all affected product in your territory and complete the Inventory Return Certification Form (Attachment 1). Email a completed copy of Attachment 1 to [email protected]. 4. Return the recalled product along with the completed Inventory Return Certification Form (Attachment 1). Clearly mark the outside carton of each product return shipment made as Recall. 5. Please notify Zimmer Biomet of any hospitals to which you have further distributed the affected product. Supply the information for any hospitals that you have identified, as well as the affected surgeons, using the provided spreadsheet template. The template will be emailed to you for completion. Please return to corporate [email protected]. 6. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-800-348-2759 between 8:00 am and 8:00 pm EST.
Quantity in commerce
384 distribution events; 1 unit was returned leaving 383 units .
Distribution
Worldwide Distribution to Germany, Belgium, Denmark, France, UK, Greece, Ireland, Netherlands, Norway, Russian Fed., Saudi Arabia, Sweden, and Thailand.
Product codes and lots
lots 62491580 (P/N 00598003701) and 62460315 (P/N 00598603701)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0590-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0590-2016Class IITerminatedVoluntary: Firm initiated