Z-0597-2016Class II

Software version VD10E for Syngo X-Workplace; Picture archiving and communication system

Siemens Medical Solutions USA, Inc. / initiated 2015-11-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Software version VD10E for Syngo X-Workplace; Picture archiving and communication system on . FDA classified it as Class II and gives the reason as "Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0597-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Software version VD10E for Syngo X-Workplace; Picture archiving and communication system.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent an Important Customer Safety Notice dated November 17, 2015. The Notice identified the product, the problem, and the action to be taken by the customer. The potential issue will be remedied with a software update via Update Instructions AX067/15/S. Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers with questions were advised to call 800-888-7436.
Quantity in commerce
73 units
Distribution
Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
Product codes and lots
Software Version: SW VD-10E

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0597-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0597-2016Class IITerminatedVoluntary: Firm initiated