510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0760-2016Class II

Arctic Sun 5000 and Arctic Sun 2000; Intended for monitoring and controlling patient temperatures

C.R. Bard Inc / initiated 2015-11-10 / Terminated

C.R. Bard, Inc. began a recall of Arctic Sun 5000 and Arctic Sun 2000; Intended for monitoring and controlling patient temperatures on . FDA classified it as Class II and gives the reason as "Improper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0760-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Arctic Sun 5000 and Arctic Sun 2000; Intended for monitoring and controlling patient temperatures.
Product code
DWJ System, Thermal Regulating
Reason for recall
Improper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe.
FDA root cause
Use error
Action
Bard Medical Division sent a Safety Alert to each of the customers via FedEx with proof of delivery notice. All consignees have been instructed to carry the notification out to the user-level.
Quantity in commerce
1,539
Distribution
Nationwide Distribution-DC and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI. IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
Product codes and lots
All lots/serial numbers sold prior to August 2014.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0760-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0760-2016Class IITerminatedVoluntary: Firm initiated