Z-0562-2016Class II

MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206 MicroScan Pos (Positive) panels are designed...

Siemens Healthcare Diagnostics Inc. / initiated 2015-12-01 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent... on . FDA classified it as Class II and gives the reason as "Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0562-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan Pos Breakpoint Combo Panel Type 23, Part No. B1017-206 MicroScan Pos (Positive) panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.
Product code
LTT Panels, Test, Susceptibility, Antimicrobial
Reason for recall
Beckman Coulter has confirmed falsely negative results for Voges-Proskauer (VP) (pale-pink/brown/colorless) with Quality Control (QC) American Type Culture Collection (ATCC) organism Staphylococcus aureus ATCC 29213 in a portion of affected lots of MicroScan Pos Combo and MicroScan Pos Breakpoint Combo panels. The expected result is positive.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 1, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to discontinue use and discard any remaining inventory of lots listed; return the attached form for product replacement; retain your inventory of the other lot numbers of the panel types, as they are not impacted by this issue; discuss content of the letter with your laboratory director/staff, retain the notification as part of your laboratory QS documentation; if you have forwarded any of the affected products to another laboratory provide them a copy of the letter, and complete and return the enclosed Response Form within 10 days via mail to: Beckman Coulter, Inc., 1584 Enterprise Blvd., West Sacramento, CA 95691, Attn: Quality Systems & Compliance or Fax: 916-374-2119 or email: [email protected]. Customers with questions are instructed to contact their Customer Support Representative at http://www.beckmancoulter.com or by phone at (800) 677-7226 in US and Canada.
Quantity in commerce
715 units total (616 units in US)
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Austria, Belgium, Brazil, Canada, Germany, Italy, Japan, Kazakhstan, Kuwait, Mexico, Myanmar, Netherlands, Norway, Panama, Poland, Portugal, Romania, Russian Federation, Spain, Taiwan, Thailand and United Kingdom.
Product codes and lots
2016-05-06 Part No. B1017-206

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0562-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0558-2016Class IITerminatedVoluntary: Firm initiated
Z-0559-2016Class IITerminatedVoluntary: Firm initiated
Z-0560-2016Class IITerminatedVoluntary: Firm initiated
Z-0561-2016Class IITerminatedVoluntary: Firm initiated
Z-0562-2016Class IITerminatedVoluntary: Firm initiated