Cournand Modified 2-PC 19G, 5.4cm or (Guidewire Introducer Needle) 20 units per case The Stryker Guidewire...
Stryker Instruments Div. of Stryker Corporation began a recall of Cournand Modified 2-PC 19G, 5.4cm or (Guidewire Introducer Needle) 20 units per case The Stryker Guidewire Introducer Needles are designed for the percutaneous... on . FDA classified it as Class II and gives the reason as "The Guidewire Introducer Needle (0910-192-000) was shipped after the documented expiration date or with minimal time until expiration. There is a potential risk of a soft tissue infection to the patient if device is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0417-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Cournand Modified 2-PC 19G, 5.4cm or (Guidewire Introducer Needle) 20 units per case The Stryker Guidewire Introducer Needles are designed for the percutaneous puncture of the anterior/posterior wall of a blood vessel. This provides an entry for a guidewire, vessel dilator, sheath, and catheter introduction.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- The Guidewire Introducer Needle (0910-192-000) was shipped after the documented expiration date or with minimal time until expiration. There is a potential risk of a soft tissue infection to the patient if device is used after expiration date.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Stryker sent an Urgent Medical Device Recall Notification letter dated November 24, 2015, to all affected customers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions regarding this recall, please contact Stryker Instruments: Monday-Friday 8am-5pm (EST) 269-389-2921
- Quantity in commerce
- 120 units
- Distribution
- Nationwide Distribution including PA and SC.
- Product codes and lots
- Model Number(s): 0910-192-000 Lot Numbers Affected 61406007 expiration date: 8/31/2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0417-2016 | Class II | Terminated | Voluntary: Firm initiated |