510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0603-2016Class II

AMSCO Warming Cabinet. Manufactured by: STERIS Corporation To raise the temperature of blankets, linens and...

STERIS Corporation / initiated 2015-11-20 / Terminated

Steris Corporation began a recall of AMSCO Warming Cabinet. Manufactured by: STERIS Corporation To raise the temperature of blankets, linens and sterile surgical IV/irrigation solutions to an... on . FDA classified it as Class II and gives the reason as "The firm discovered that information printed on the quick reference magnet is incorrect; specifically, Fahrenheit to Celsius temperature conversions and the default Warming Cabinet temperature". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0603-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
AMSCO Warming Cabinet. Manufactured by: STERIS Corporation To raise the temperature of blankets, linens and sterile surgical IV/irrigation solutions to an acceptable level for various surgical, obstetrical, emergency, critical care and other healthcare applications.
Product code
LGZ Warmer, Thermal, Infusion Fluid
Reason for recall
The firm discovered that information printed on the quick reference magnet is incorrect; specifically, Fahrenheit to Celsius temperature conversions and the default Warming Cabinet temperature.
FDA root cause
Process change control
Action
The firm, Steris, sent an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated 12/14/2015 to its customers. The letter described the product, problem and actions to be taken. The firm included an updated quick reference magnet for each affected unit with the letter. The customers were instructed to replace your existing magnet with the enclosed magnet. If you have questions regarding this matter, please contact Product Manager, at 440-392-7705 or STERIS Customer Service at 1-800-548-4873.
Quantity in commerce
2,319 units
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Bahrain, Canada, China, Colombia, Costa Rica, France, Hong Kong, India, Indonesia, Israel, Kuwait, Lebanon, Mexico, Panama, Peru, Republic of Korea, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, Uganda, United Arab Emirates, Viet Nam & Virgin Islands.
Product codes and lots
Serial #'s: 0428014026 - 0432115014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0603-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0603-2016Class IITerminatedVoluntary: Firm initiated