Columbus Revision Knee System, EnduRo Knee System
Aesculap, Inc. began a recall of Columbus Revision Knee System, EnduRo Knee System Product Usage: The Columbus Revision Knee System and EnduRo Knee System are indicated for use in... on . FDA classified it as Class II and gives the reason as "Aesculap Inc. US has initiated a recall on Tibial and Femur extension sterile pressfit stem implant packaging that are used for Columbus Revision/Enduro Knee Implants because it is labeled as Cementless. US product is...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1043-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- Columbus Revision Knee System, EnduRo Knee System Product Usage: The Columbus Revision Knee System and EnduRo Knee System are indicated for use in reconstruction of the diseased knee joint.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Aesculap Inc. US has initiated a recall on Tibial and Femur extension sterile pressfit stem implant packaging that are used for Columbus Revision/Enduro Knee Implants because it is labeled as Cementless. US product is indicated in the US for use with Bone Cement only.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Aesculap sent an Urgent Medical Device Correction letter to Distributors and Sales Rep. The letter identified the affected product, problem and actions to be taken. Customers were notified to add the over label to product packaging. Telephone conferences with Distributors and Sales Rep reiterated need for representative to inform Surgical Handler of the label when providing product for review. Customers were instructed to complete the attached Product Correction Acknowledgement Form and return to Aesculap Quality Assurance department by faxing the form to 610-791-6882 or e-mail to [email protected], two (2) weeks of receipt, even if the total inventory in your possession is zero (0).
- Quantity in commerce
- 1232
- Distribution
- US Nationwide Distribution
- Product codes and lots
- All codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1043-2016 | Class II | Terminated | Voluntary: Firm initiated |