Z-0467-2016Class II

MAGNETOM systems, magnetic resonance diagnostic devices (MRDD)

Siemens Medical Solutions USA, Inc. / initiated 2015-12-07 / Terminated

Siemens Medical Solutions USA, Inc began a recall of MAGNETOM systems, magnetic resonance diagnostic devices (MRDD) on . FDA classified it as Class II and gives the reason as "Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0467-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.
FDA root cause
Other
Action
Siemens sent an Customer Advisory letter dated December 7, 2015, certified mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter notifies customers of the potential injuries, which may result of improperly installed quench lines. The letter indicates it is the system's owner who is responsible for installing the quench line per the requirements. If Siemens installed the quench lines, the consignee is directed to contact Siemens, who will inspect the quench lines during next maintenance visit. Siemens will monitor service reports for affected sites. For further questions please call ( 610) 219-6300.
Quantity in commerce
in US 3752 units
Distribution
Nationwide including : DC, PR, Virgin Islands and Bermuda
Product codes and lots
all units distributed since 1991

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0467-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0467-2016Class IITerminatedVoluntary: Firm initiated