Z-0467-2016Class II
MAGNETOM systems, magnetic resonance diagnostic devices (MRDD)
Siemens Medical Solutions USA, Inc. / initiated 2015-12-07 / Terminated
Siemens Medical Solutions USA, Inc began a recall of MAGNETOM systems, magnetic resonance diagnostic devices (MRDD) on . FDA classified it as Class II and gives the reason as "Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0467-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.
- FDA root cause
- Other
- Action
- Siemens sent an Customer Advisory letter dated December 7, 2015, certified mail to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter notifies customers of the potential injuries, which may result of improperly installed quench lines. The letter indicates it is the system's owner who is responsible for installing the quench line per the requirements. If Siemens installed the quench lines, the consignee is directed to contact Siemens, who will inspect the quench lines during next maintenance visit. Siemens will monitor service reports for affected sites. For further questions please call ( 610) 219-6300.
- Quantity in commerce
- in US 3752 units
- Distribution
- Nationwide including : DC, PR, Virgin Islands and Bermuda
- Product codes and lots
- all units distributed since 1991
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0467-2016 | Class II | Terminated | Voluntary: Firm initiated |