Z-0701-2016Class II

Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit...

Datascope Corp. / initiated 2015-12-15 / Terminated / root cause: software

Mindray DS USA, Inc. dba Mindray North America began a recall of Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A... on . FDA classified it as Class II and gives the reason as "An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0701-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Panorama Patient Monitoring Network. Software Kit 0020-00-0205-19A, View Station Software Kit 0020-00-0206-16A, Work Station Software Kit 0020-00-0207-16A, Software Kit Upgrade TECH-00-910. The Panaroma Patient Monitoring Network is intended for use in a fixed location in the healthcare facility setting as a central viewing station. It is not intended to be directly connected to patient at any time or installed in a patient's vicinity.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
An issue with the Panorama Central Station may cause the system to spontaneously restart. This may occur after about 49 days of continuous operation. On restart, monitoring will be lost for about 2.5 minutes. Settings and configurations are maintained.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Mindray DS USA, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated 12/15/2015 to its customers via certified mail. The letter described the product, problem, and actions to be taken. The customers were instructed to route the recall letter to all potential users of your facility. Mindray will resolve the issue by upgrading the system software on all affected Panorama systems in your facility. Please contact your Mindray Service Representative at 1-800-288-2121 (Monday-Friday 8:30am-5:30pm) to arrange upgrade. If you have any questions, please contact Director, Quality Operations and Regulatory Affair at 201-995-8407 or email to: [email protected].
Quantity in commerce
63 products
Distribution
Nationwide Distribution.
Product codes and lots
version 11.5.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0701-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0701-2016Class IITerminatedVoluntary: Firm initiated