arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium...
Arkray Factory USA, Inc. began a recall of arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin... on . FDA classified it as Class I and gives the reason as "One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0631-2016
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arkray Factory USA, Inc.
- Product
- arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The firm sent an Urgent Medical Device Recall letter to customers beginning 12/18/2015. The letter identified affected product, stated the reason for recall and asked for unused product to be returned. For instructions on returning unused product contact Lesty Jaranilla 1-877-538-8872 x3243.
- Quantity in commerce
- 11 boxes (25 foil packaged test strips per box)
- Distribution
- Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
- Product codes and lots
- Lot Number PN5C26
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0630-2016 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0631-2016 | Class I | Terminated | Voluntary: Firm initiated |