Z-0604-2016Class II

Oncentra External Beam - VMAT - Radiation therapy planning system

Elekta Limited / initiated 2015-12-18 / Terminated / root cause: software

Elekta, Inc. began a recall of Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to... on . FDA classified it as Class II and gives the reason as "When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0604-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.
Root cause tag
software
FDA root cause
Software design
Action
Elekta sent an Important Field Safety Notice 806-01-ETP-001 to all affected customers on December 18, 2015. The notice informs users of the issue and that should the issue occur it could result in overdosage of organs at risk outside the intended treatment volume. Users are strongly advised to perform proper Quality Assurance for all treatment plans before delivery of the first fraction to the patient. To prevent the issue from occurring, the workaround is to not use the Tumor Overlap Fraction for VMAT planning.and not to use the Tumor Overlap Fraction for VMAT planning. The issue will be solved in Oncentra External Beam version 4.5.2. The notice included an acknowledgement form which is to be returned to Elekta. For questions contact your local Elekta representative.
Quantity in commerce
154
Distribution
Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.
Product codes and lots
Versions 3.3 and higher

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0604-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0604-2016Class IITerminatedVoluntary: Firm initiated