BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Tan) BD Hemogard Additive: K2EDTA (spray...

Becton Dickinson & Co / initiated 2015-12-03 / Terminated / root cause: manufacturing process

Becton Dickinson & Company began a recall of BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Tan) BD Hemogard Additive: K2EDTA (spray dried), 5.4mg Sterile IVD BD, Made in USA BD... on . FDA classified it as Class II and gives the reason as "Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0570-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Tan) BD Hemogard Additive: K2EDTA (spray dried), 5.4mg Sterile IVD BD, Made in USA BD Vacutainer tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory. The BD Vacutainer Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
The firm, BD, issued an "URGENT PRODUCT RECALL" notifications/Recall Response Forms to their customers dated 12/3/2015 via email and/or UPS 2nd Day Mail. The notification identified the reason for the recall; how to identify affected product; and actions to be taken. The customers were instructed customers to immediately review their inventory; complete the enclosed Recall Response Form and fax (1-866-873-0312 it to BD or email it to [email protected]; and return all affected products with the completed Recall Response Form following instructions on the enclosed packing instruction, even if you do not have any affected lot in your inventory. If customers have any questions or require assistance with the return of the recalled product and/or availability of replacement product, they were instructed to contact 1-888-628-0732 between 8AM and 5 PM ET Monday through Friday.
Quantity in commerce
29,815,600 units in total
Distribution
Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
Product codes and lots
Catalog Reference #367855; Lot numbers 5064684, 5120653, 5154643, 5175930, 5210765, 5237787

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0570-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0567-2016Class IITerminatedVoluntary: Firm initiated
Z-0568-2016Class IITerminatedVoluntary: Firm initiated
Z-0569-2016Class IITerminatedVoluntary: Firm initiated
Z-0570-2016Class IITerminatedVoluntary: Firm initiated
Z-0571-2016Class IITerminatedVoluntary: Firm initiated