510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0628-2016Class III

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent

Becton, Dickinson and Company, BD Biosciences / initiated 2015-12-14 / Terminated

Becton, Dickinson and Company, BD Biosciences began a recall of FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent on . FDA classified it as Class III and gives the reason as "Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0628-2016
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton, Dickinson and Company, BD Biosciences
Product
FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
Product code
MVU Reagents, Specific, Analyte
Reason for recall
Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.
FDA root cause
Under Investigation by firm
Action
BD Biosciences sent an Urgent Product Recall letter dated December 2015, to all affected customers to inform them that Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to return the Tracking/Verification form to BD Biosciences. Customers are instructed to contact BD Customer Support at (877) 232-8995 (prompt 3, 3) in the US. Customers outside of the US are instructed to contact their local BD Biosciences or distributor.
Quantity in commerce
1,388 units
Distribution
Nationwide Distribution
Product codes and lots
Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0628-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0628-2016Class IIITerminatedVoluntary: Firm initiated