FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent
Becton, Dickinson and Company, BD Biosciences began a recall of FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent on . FDA classified it as Class III and gives the reason as "Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0628-2016
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
- Product code
- MVU Reagents, Specific, Analyte
- Reason for recall
- Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.
- FDA root cause
- Under Investigation by firm
- Action
- BD Biosciences sent an Urgent Product Recall letter dated December 2015, to all affected customers to inform them that Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to return the Tracking/Verification form to BD Biosciences. Customers are instructed to contact BD Customer Support at (877) 232-8995 (prompt 3, 3) in the US. Customers outside of the US are instructed to contact their local BD Biosciences or distributor.
- Quantity in commerce
- 1,388 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0628-2016 | Class III | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
