Z-1248-2016Class II
Cardiosave Hybrid and rescue IABP
Datascope Corp. / initiated 2015-12-16 / Terminated
Maquet Datascope Corp - Cardiac Assist Division began a recall of Cardiosave Hybrid and rescue IABP on . FDA classified it as Class II and gives the reason as "Maquet has recieved information that in some Cardiosave IABPS, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1248-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Cardiosave Hybrid and rescue IABP
- Product code
- DSP System, Balloon, Intra-Aortic And Control
- Reason for recall
- Maquet has recieved information that in some Cardiosave IABPS, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates.
- FDA root cause
- Other
- Action
- Maquet sent an "Urgent - Medical Device Field Correction Notice and Response form dated December 16, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are informed of the issues and provided corrective actions. Customers are asked to completed the attached response form in order to return to 1-973-629-1518, or via email to [email protected]. Questions can be directed 1-800-777-4222, option 3, Monday through Friday 8 am-6 pm. EST.
- Quantity in commerce
- 1243 units US; 1107 units OUS
- Distribution
- Worldwide Distribution- Nationwide Distribution including Puerto Rico.
- Product codes and lots
- Cardiosave Hybrid IABP 0998-00-0800-53,0998-UC-0800-53, Cardiosave Rescue IABP 0998-00-0800-83 & 0998-UC-0800-83
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1248-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY