Z-0659-2016Class II

VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the...

LivaNova USA, Inc. / initiated 2015-12-18 / Terminated / root cause: design

Cyberonics, Inc began a recall of VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures on . FDA classified it as Class II and gives the reason as "Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0659-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.
Product code
LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
Reason for recall
Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.
Root cause tag
design
FDA root cause
Device Design
Action
The firm notified consignees of the issue via letter on 12/18/15. The letter identified the affected device, the issue involved, and actions to be taken. Physicians are to contact Clinical Technical Support at 866-882-8804 to report if a patient's generator has been disabled due to the issue identified. Users are to complete and return the effectiveness card as soon as possible. If further information is needed, customers can contact Clinical Technical Support at 866-882-8804 or via e-mail at [email protected].
Quantity in commerce
4,935 units
Distribution
Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.
Product codes and lots
All VNS Therapy AspireSR (Model 106) Generators; Device Identifier - (01)05425025750061

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0659-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0659-2016Class IITerminatedVoluntary: Firm initiated