510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0818-2016Class II

High Flow Irrigation Extender; Part Number 72201021. Arthroscopic accessories are indicated for use in access...

Smith & Nephew, Inc. / initiated 2013-07-30 / Terminated

Smith & Nephew, Inc., Endoscopy Div. began a recall of High Flow Irrigation Extender; Part Number 72201021. Arthroscopic accessories are indicated for use in access of the joint capsule on . FDA classified it as Class II and gives the reason as "During arthroscopic surgery, when the scope is moved inside the cannula, the body of the irrigation extender separates, potentially obstructing visualization and causing leakage of the irrigation fluid. The device is...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0818-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
High Flow Irrigation Extender; Part Number 72201021. Arthroscopic accessories are indicated for use in access of the joint capsule.
Product code
NBH Accessories, Arthroscopic
Reason for recall
During arthroscopic surgery, when the scope is moved inside the cannula, the body of the irrigation extender separates, potentially obstructing visualization and causing leakage of the irrigation fluid. The device is removed and a back-up device is used to complete the surgery.
FDA root cause
Under Investigation by firm
Action
Urgent - Product Recall 1st Notification - Urgent R-2013-13 letters, dated July 30, 2013 were sent overnight via Federal Express to customers. The letters identified the affected product, as well as the potential risk involved. Actions to be taken were provided for both hospital and sales representatives. Hospital representatives were asked to inspect their inventory, locate all devices from the list provided, and quarantine them immediately. The Inventory Return Certification Form should be completed and returned, regardless of whether any affected product is on hand or not. Contact the Smith & Nephew Returns Group at 800-343-5717 option 3) or send an e-mail to [email protected]. They will provide instructions for returning affected product and receiving a replacement.
Quantity in commerce
203 units
Distribution
Worldwide Distribution -- US, including the states of AL, AR, CA, FL, GA, ID, IL, KS, MA, MI, MO, NC, TN, TX, and AK; and, country of Japan.
Product codes and lots
50209581 50220887 50231401 50216929 50226220 50231932 50218285 50229266 50235063 50240019 50245683 50251109 50254562 50256942 50263784 50279382 50291593 50302984 50303418 50327069 50330377 50351345 50361381 50404796 50408768 50419699 50433964 50250837 50256854 50269580 5027766 50304642 50342002 50401593 50417254

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0818-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0818-2016Class IITerminatedVoluntary: Firm initiated