Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2015-12-15 / Terminated / root cause: design

Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111 on . FDA classified it as Class II and gives the reason as "The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0693-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).
Root cause tag
design
FDA root cause
Device Design
Action
Fresenius Medical Care issued on December 15, 2015, a n Important Product Notification to all affected customers informing them of the problem and how to recognize the proper latch closing. Questions regarding the contents of the Customer Notification or questions regarding the issue, contact Technical Services at 1-800-227-2572 and reference the Field Action Number FA-2015-10-W. A reply form is included as part of the customer communication with instructions to fill out and return via fax or scanned/emailed copy.
Quantity in commerce
14,094 units
Distribution
Nationwide Distribution
Product codes and lots
All lot numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0693-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0693-2016Class IITerminatedVoluntary: Firm initiated