Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111
Fresenius Medical Care Renal Therapies Group, LLC began a recall of Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111 on . FDA classified it as Class II and gives the reason as "The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0693-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111
- Product code
- FKX System, Peritoneal, Automatic Delivery
- Reason for recall
- The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Fresenius Medical Care issued on December 15, 2015, a n Important Product Notification to all affected customers informing them of the problem and how to recognize the proper latch closing. Questions regarding the contents of the Customer Notification or questions regarding the issue, contact Technical Services at 1-800-227-2572 and reference the Field Action Number FA-2015-10-W. A reply form is included as part of the customer communication with instructions to fill out and return via fax or scanned/emailed copy.
- Quantity in commerce
- 14,094 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- All lot numbers
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0693-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY