510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0654-2016Class II

Zimmer Universal Locking System 3.5mm Locking Reconstruction Plate Straight 11 Holes 144mm Length Zimmer...

Zimmer GmbH / initiated 2015-12-09 / Terminated / root cause: labeling and UDI

Zimmer Biomet, Inc. began a recall of Zimmer Universal Locking System 3.5mm Locking Reconstruction Plate Straight 11 Holes 144mm Length Zimmer Universal Locking System 3.5mm Locking Dual... on . FDA classified it as Class II and gives the reason as "The product on the package label did not match what was inside the package. The label was for part number 00-4936-011-13 lot number 62968956 while the product inside the package was part number 00-4936-008-07 lot number...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0654-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer Universal Locking System 3.5mm Locking Reconstruction Plate Straight 11 Holes 144mm Length Zimmer Universal Locking System 3.5mm Locking Dual Compression Plate 8 Holes 105mm Length. Indicated for temporary internal fixation and stabilization of osteotomies and fractures.
Product code
HRS Plate, Fixation, Bone
Reason for recall
The product on the package label did not match what was inside the package. The label was for part number 00-4936-011-13 lot number 62968956 while the product inside the package was part number 00-4936-008-07 lot number 62968974.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On 12/9/2015, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital risk managers via courier with instructions for identifying and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Specifically, customers were asked to review the notification and ensure all affected personnel are aware of the information; assist the Zimmer Biomet sales representative with the quarantine and removal of any of the identified affected product; and complete and return the Certificate of Acknowledgement form. If after reviewing this notification customers have further questions or concerns please call the customer call center at 1-877-946-2761. Hours of operation are Monday through Friday, 8 a.m. through 5 p.m. EST.
Quantity in commerce
52
Distribution
Worldwide Distribution -- USA, including the states of MN, NY, MS, AL, OH, KY, ME, and NM; and, the countries of Taiwan, Korea, Singapore, El Salvador, Uruguay, Czech Republic, France, Italy, Poland, United Kingdom, South Africa, and Slovakia.
Product codes and lots
Part 00-4936-011-13, Lot 62968956 3.5MM ULS RECON PLATE 11H Part 00-4936-008-07, Lot 62968974 3.5MM ULS DUAL COMP PLT 8

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0654-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0654-2016Class IITerminatedVoluntary: Firm initiated