Z-0737-2016Class II
Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters
Bard Peripheral Vascular Inc / initiated 2015-03-13 / Terminated / root cause: labeling and UDI
Bard Peripheral Vascular Inc began a recall of Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters on . FDA classified it as Class II and gives the reason as "IFU missing contraindications: Patients with uncontrolled sepsis and Patients with known hypersensitivity to nickel-titanium alloys". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0737-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters.
- Product code
- DTK Filter, Intravascular, Cardiovascular
- Reason for recall
- IFU missing contraindications: Patients with uncontrolled sepsis and Patients with known hypersensitivity to nickel-titanium alloys.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- The firm sent out Medical Device Recall customer notification letters on 03/13/15. The notification letter stated that there is no required action on the customer's part. Customers were told that if the device did not meet their needs then the firm will provide a replacement unit. Effective immediately the firm is implementing a change in the ordering code for the device to accurately reflect the approved information with the IFU and corresponding retrieval kit. Any questions the firm states to call Bard Canada: 1-800-268-2862 Western Canada 1-800-387-9473 Ontario 1-800-387-7851 Eastern Canada
- Quantity in commerce
- 1,183 units
- Distribution
- Distributed to CA only.
- Product codes and lots
- Product Code: DL950J Lot Number: GFYA3093, GFYC3745, GFYD2734, GFYF3878, GFYF3879, GFYG3877, GFYG3880, GFYI2699 Product Code: DL950F Lot Number: GFYA3093, GFYC3745, GFYD2734, GFYF3878, GFYF3879, GFYG3877, GFYG3880, GFYI2699