510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0737-2016Class II

Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters

Bard Peripheral Vascular Inc / initiated 2015-03-13 / Terminated / root cause: labeling and UDI

Bard Peripheral Vascular Inc began a recall of Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters on . FDA classified it as Class II and gives the reason as "IFU missing contraindications: Patients with uncontrolled sepsis and Patients with known hypersensitivity to nickel-titanium alloys". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0737-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters.
Product code
DTK Filter, Intravascular, Cardiovascular
Reason for recall
IFU missing contraindications: Patients with uncontrolled sepsis and Patients with known hypersensitivity to nickel-titanium alloys.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The firm sent out Medical Device Recall customer notification letters on 03/13/15. The notification letter stated that there is no required action on the customer's part. Customers were told that if the device did not meet their needs then the firm will provide a replacement unit. Effective immediately the firm is implementing a change in the ordering code for the device to accurately reflect the approved information with the IFU and corresponding retrieval kit. Any questions the firm states to call Bard Canada: 1-800-268-2862 Western Canada 1-800-387-9473 Ontario 1-800-387-7851 Eastern Canada
Quantity in commerce
1,183 units
Distribution
Distributed to CA only.
Product codes and lots
Product Code: DL950J Lot Number: GFYA3093, GFYC3745, GFYD2734, GFYF3878, GFYF3879, GFYG3877, GFYG3880, GFYI2699  Product Code: DL950F Lot Number: GFYA3093, GFYC3745, GFYD2734, GFYF3878, GFYF3879, GFYG3877, GFYG3880, GFYI2699

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0737-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0737-2016Class IITerminatedVoluntary: Firm initiated