Z-0661-2016Class II

Fukuda Denshi patient monitor model DS-8100M and DS-8100N. The Fukuda Denshi DynaScope Model DS-8100N/8100M...

Fukuda Denshi USA, Inc. / initiated 2015-12-21 / Terminated / root cause: manufacturing process

Fukuda Denshi USA, Inc. began a recall of Fukuda Denshi patient monitor model DS-8100M and DS-8100N. The Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is meant to acquire and monitor... on . FDA classified it as Class II and gives the reason as "A possible malfunction where a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term Backup battery)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0661-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fukuda Denshi USA, Inc.
Product
Fukuda Denshi patient monitor model DS-8100M and DS-8100N. The Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is meant to acquire and monitor physiological signals from patients.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
A possible malfunction where a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term Backup battery).
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Fukuda Denshi USA, Inc. of Redmond, WA plans to perform 100% field correction of units distributed in the US. The Recall/Field Correction Notice letter (FCN-A01-201512), dated December 21, 2015, and the Customer Confirmation Sheet will be sent to consignees using a clearly red-marked URGENT via Express Mail. The consignees are requested to fill out the Customer Confirmation Sheet and return via fax to 888-224-7090, email to [email protected], or send via postal service back to Fukuda Denshi USA, Inc. in a pre-stamped envelope. Consignee can call Doug Blakey directly at 210-240-4521 with any questions or concerns.
Quantity in commerce
95 units distributed in the US
Distribution
Distributed in the states of CO, ID, KS, MI, MO, OR, PA, TX, UT, and WI.
Product codes and lots
50400001, 50400002, 50400003, 50400004, 50400011, 50400012, 50400018, 50400019, 50400020, 50400021, 50400022, 50400023, 50400024, 50400025, 50400026, 50400027, 50400028, 50400029, 50400030, 50400031, 50400032, 50400033, 50400034, 50400035, 50400036, 50400037, 50400038, 50400039, 50400040, 50400042, 50400043, 50400044, 50400045, 50400046, 50400047, 50400048, 50400049, 50400050, 50400051, 50400052, 50400053, 50400054, 50400055, 50400056, 50400057, 50400058, 50400059, 50400060, 50400061, 50400062, 50400063, 50400064, 50400065, 50400066, 50400067, 50400068, 50400069, 50400070, 50400071, 50400072, 50400073, 50400074, 50400075, 50400076, 50400077, 50400078, 50400084, 50400093, 50400094, 50400095, 50400096, 50400097, 50400098, 50400099, 50400100, 50400101, 50400102, 50400103, 50400104, 50400105, 50400106, 50400107, 50400108, 50400109, 50400110, 50400111, 50400112, 50400113, 50400114, 50400115, 50400116, 50400117, 50400118, 50400119, and 50400120.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0661-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0661-2016Class IITerminatedVoluntary: Firm initiated