Z-1345-2016Class II

NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No...

Boston Scientific Corporation / initiated 2015-11-03 / Terminated / root cause: manufacturing process

AngioDynamics Inc. began a recall of NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This needle is intended... on . FDA classified it as Class II and gives the reason as "Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1345-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
NAMIC CONVENIENCE KIT HOLYOKE HOSPITAL SPECIAL PROCEDURE KIT, REF Catalog No. 60080094, UPN Product No. H749600800941 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire
Product code
OEZ Cardiovascular Procedure Kit
Reason for recall
Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On November 03, 2015, Navilyst Medical distributed Urgent Voluntary Medical Device Recall notices to their customers via courier service. Customers were advised to immediately remove the recall product from inventory and to segregate the products in a secure location for return to AngioDynamics, Inc. Customers should immediately forward a copy of the recall notification letter to all sites if these products were further distributed. If affected product is located in your institution, please call AngioDynamics, Inc. Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday EST) to obtain a replacement or credit for your returned product. Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return). The Reply Verification Tracking form can be submitted via fax to: Attn: Convenience Kits - Procedure Products Needles 1-800-782-1357 or via email to [email protected] or [email protected].
Quantity in commerce
8 kits
Distribution
US Nationwide in the states of CA, IL, MA, SC, TX
Product codes and lots
Batch/Lo. 4925662 Use By 2017-02

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1345-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1343-2016Class IITerminatedVoluntary: Firm initiated
Z-1344-2016Class IITerminatedVoluntary: Firm initiated
Z-1345-2016Class IITerminatedVoluntary: Firm initiated
Z-1346-2016Class IITerminatedVoluntary: Firm initiated