UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for...
OrthoPediatrics Corp began a recall of UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior... on . FDA classified it as Class II and gives the reason as "Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0622-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthopediatrics Corp
- Product
- UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- The firm, OrthoPediatrics Corp., sent an "URGENT MEDICAL DEVICE RECALL" letter dated 11/13/2015 to affected customers via email. The letter included a description of the reason for the recall, affected product, consignee responsibilities, and actions to be taken. The customers were instructed to take immediate possession of all the identified devices; quarantine them and immediately return them to OrthoPediatrics, and complete and return the Mandatory Reply Form via fax to (574) 269-3692 or email: [email protected]. If you have any questions, please contact VP-Regulatory Affairs (574) 268-6379 or email: [email protected].
- Quantity in commerce
- 102
- Distribution
- US Distribution to states of: AL, AZ, CA, DE, FL, GA, HI, KY, LA, MI, MO, MN, MS, NC, NJ, NY, OH, OR, TN, and TX.
- Product codes and lots
- Item number; lot; product description 00-1300-1735; 44003; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 35MM. 00-1300-0735; 44601; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 35MM. 00-1300-0740; 44602; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 43926; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 44252; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0745; 43927; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 45MM. 00-1300-0750; 44619; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 50MM. 00-1300-0620; 44623; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 20MM. 00-1300-0640; 44617; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-0640; 44884; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-1745; 44005; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 45MM. 00-1300-0635; 44677; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 35MM.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0622-2016 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.