Z-0622-2016Class II

UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for...

Orthopediatrics Corp / initiated 2015-11-10 / Terminated / root cause: manufacturing process

OrthoPediatrics Corp began a recall of UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior... on . FDA classified it as Class II and gives the reason as "Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0622-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthopediatrics Corp
Product
UniAxial Pedicle Screw & UniAxial Reduction Pedicle Screw The Response 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Response Spine UniAxial Pedicle & UniAxial Reduction Pedicle Screws may have been manufactured incorrectly, thus preventing the driver from fully engaging into the hexagonal drive feature of the screw.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
The firm, OrthoPediatrics Corp., sent an "URGENT MEDICAL DEVICE RECALL" letter dated 11/13/2015 to affected customers via email. The letter included a description of the reason for the recall, affected product, consignee responsibilities, and actions to be taken. The customers were instructed to take immediate possession of all the identified devices; quarantine them and immediately return them to OrthoPediatrics, and complete and return the Mandatory Reply Form via fax to (574) 269-3692 or email: [email protected]. If you have any questions, please contact VP-Regulatory Affairs (574) 268-6379 or email: [email protected].
Quantity in commerce
102
Distribution
US Distribution to states of: AL, AZ, CA, DE, FL, GA, HI, KY, LA, MI, MO, MN, MS, NC, NJ, NY, OH, OR, TN, and TX.
Product codes and lots
Item number; lot; product description  00-1300-1735; 44003; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 35MM. 00-1300-0735; 44601; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 35MM. 00-1300-0740; 44602; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 43926; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0740; 44252; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 40MM. 00-1300-0745; 43927; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 45MM. 00-1300-0750; 44619; 5.5/6.0 UNIAXIAL PEDICLE SCREW 7.0MM X 50MM. 00-1300-0620; 44623; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 20MM. 00-1300-0640; 44617; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-0640; 44884; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 40MM. 00-1300-1745; 44005; 5.5/6.0 UNIAXIAL REDUCTION PEDICLE SCREW 7.0MM X 45MM. 00-1300-0635; 44677; 5.5/6.0 UNIAXIAL PEDICLE SCREW 6.0MM X 35MM.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0622-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0622-2016Class IITerminatedVoluntary: Firm initiated