510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0785-2016Class II

Bard Distaflo Vascular Bypass Graft

Bard Peripheral Vascular Inc / initiated 2016-01-05 / Terminated / root cause: labeling and UDI

Bard Peripheral Vascular Inc began a recall of Bard Distaflo Vascular Bypass Graft; Product Code: DF6006SC; 8 mm x 60 cm. Intended for bypass or reconstruction of peripheral arterial blood vessels on . FDA classified it as Class II and gives the reason as "Size mislabeling; the graft is labeled as a 6 mm diameter graft when the actual packaged product is an 8 mm graft". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0785-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
Bard Distaflo Vascular Bypass Graft; Product Code: DF6006SC; 8 mm x 60 cm. Intended for bypass or reconstruction of peripheral arterial blood vessels.
Product code
DSY Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Reason for recall
Size mislabeling; the graft is labeled as a 6 mm diameter graft when the actual packaged product is an 8 mm graft.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Customer notification letters were sent on 01/05/2016. Customers were asked to not use the affected product, or further distribute it. Customers were asked to check all inventory locations for the affected product code, lot number combination listed. If customers further distributed any of the product they are requested to contact that location, advise them of the recall, and forward the listed instructions. The letter states to remove all identified product from the shelves. If no affected product is going to be returned the firm states to complete and return the attached form stating so. If customers have product it is requested that they also complete the attached form, and state the quantities and lot numbers in stock. The letter requests that customers call the firm's Recall Coordinator, Raye Seisinger, at 1-800-321-4254, option #2 ext. 2501 (M-F 6am to 3pm MST) or email [email protected]. Once all information has been verified the customer will be issued a Return Authorization or Consignment Recall Number to facilitate the expedient return of the product. The firm will provide replace product for returned product. Fax the attached form to the firm at 1-800-994-6772. If the customer cannot fax the firm they request that the Recall Coordinator be contact and provided the information verbally. The letter states a mailing label is enclosed to return the affected product. Customers are requested to amek the package as "recalled product" and include the XC or XH number. All products should be returned to the following shipping address: Bard Peripheral Vascular, Inc. 1415 W. 3rd St. Tempe, AZ 85281
Quantity in commerce
4 units
Distribution
Distributed in the states of AL, TX, and CA.
Product codes and lots
Product Code: DF6006SC Lot Number: VTZE1782

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0785-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0785-2016Class IITerminatedVoluntary: Firm initiated
Z-0786-2016Class IITerminatedVoluntary: Firm initiated