510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0745-2016Class II

DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System...

Philips North America LLC / initiated 2016-01-08 / Terminated / root cause: design

Philips Medical Systems began a recall of DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #459800200621 on . FDA classified it as Class II and gives the reason as "In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0745-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #459800200621
Product code
IYE Accelerator, Linear, Medical
Reason for recall
In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.
Root cause tag
design
FDA root cause
Device Design
Action
Philips sent an "Urgent Medical Device Correction" letter dated 16-DEC-2015 to affected customers. The letter described the Affected Products, Problem Description, Hazard Involved, How to Identify Affected Products, Action to be Taken by Customer / User, Actions Planned by Philips and Further Information and Support (contact Philips Customer Care Solutions Center at 1-800-722-9377 ).
Quantity in commerce
3
Distribution
US: Nationwide Distribution in the states of FL, OH, and MO.
Product codes and lots
Software Version 10.0.0 and 14.0.0

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0745-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0745-2016Class IITerminatedVoluntary: Firm initiated