Z-0687-2016Class II

The e.cam Dual Signature Gamma Camera system. Used to detect or image the distribution of radionuclides in...

Siemens Medical Solutions USA, Inc. / initiated 2015-12-31 / Terminated / root cause: packaging

Siemens Medical Solutions USA, Inc. began a recall of The e.cam Dual Signature Gamma Camera system. Used to detect or image the distribution of radionuclides in the body or organ, using the following techniques... on . FDA classified it as Class II and gives the reason as "To provide operator instruction manuals for the e.cam Dual Signature; Gamma Camera that were not sent in its entirety when the e.cam systems were refurbished and shipped to the customer". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0687-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
The e.cam Dual Signature Gamma Camera system. Used to detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 588keV.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
To provide operator instruction manuals for the e.cam Dual Signature; Gamma Camera that were not sent in its entirety when the e.cam systems were refurbished and shipped to the customer.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
An updated Customer Advisory Notice Letter will be provided to each impacted consignee. The receipt of the letter will be documented and additional attempts will be made until such receipt is acknowledged. The Customer Advisory Notice asks that the letter be placed with systems instructions for use along with all provided user documentation. It also advises the consignee to forward the notice to the new customer if the equipment was sold. Customers with questions regarding the Customer Advisory Notice can contact their local Service representative or use the following contact number 1-800-888-7436.
Quantity in commerce
10
Distribution
Distributed in the states of Florida, New York, Indiana, Maryland, South Dakota, Missouri, Tennessee, and Georgia, and the countries of Australia and India.
Product codes and lots
e.cam systems that were refurbished from 04/06/15 to 10/06/15. Material number 10151531, serial number 70004; Material number 10151532 serial numbers 70038, 70040, 70041, 70042, 70043, 70044, 70039, 70045, and 70046.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0687-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0687-2016Class IITerminatedVoluntary: Firm initiated