Z-0862-2016Class II

Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process...

Siemens Medical Solutions USA, Inc. / initiated 2015-12-22 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process, read, report, communicate, distribute, store and... on . FDA classified it as Class II and gives the reason as "Siemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0862-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process, read, report, communicate, distribute, store and archive digital medical images. Stores and archives within DICOM structured reports. Integrates hospital/radiology information systems.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Siemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading studies; breast region is now properly fitted to segment boundary when clicking fit breast to screen.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Voluntary Correction and Removal Action letter dated December 30, 2015. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that a software update VB10A_HF01/06 is being released simulataneously to eliminate the issue. Customers with questions were instructed to call 1-888-826-9702.
Quantity in commerce
68
Distribution
Nationwide Distribution including AL CA CT FL IN KS LA MA MD MI MN MO MS NE NY OH OR PA TX WA WI
Product codes and lots
Device model numbers: 10863171, 10863172, 10863173

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0862-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0862-2016Class IITerminatedVoluntary: Firm initiated