Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process...
Siemens Medical Solutions USA, Inc began a recall of Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process, read, report, communicate, distribute, store and... on . FDA classified it as Class II and gives the reason as "Siemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0862-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo Plaza Picture archiving and communication system. Software only. PACS intended to display, process, read, report, communicate, distribute, store and archive digital medical images. Stores and archives within DICOM structured reports. Integrates hospital/radiology information systems.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Siemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading studies; breast region is now properly fitted to segment boundary when clicking fit breast to screen.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a Voluntary Correction and Removal Action letter dated December 30, 2015. The letter identified the product, the problem, and the action to be taken by the customer. Customers were notified that a software update VB10A_HF01/06 is being released simulataneously to eliminate the issue. Customers with questions were instructed to call 1-888-826-9702.
- Quantity in commerce
- 68
- Distribution
- Nationwide Distribution including AL CA CT FL IN KS LA MA MD MI MN MO MS NE NY OH OR PA TX WA WI
- Product codes and lots
- Device model numbers: 10863171, 10863172, 10863173
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0862-2016 | Class II | Terminated | Voluntary: Firm initiated |