Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures
Stryker Howmedica Osteonics Corp. began a recall of Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures on . FDA classified it as Class II and gives the reason as "The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be resterilized. Since the IFU...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0855-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures.
- Product code
- JDR Staple, Fixation, Bone
- Reason for recall
- The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be resterilized. Since the IFU states the devices are sterile, no instructions for moist heat sterilization are provided.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Stryker Branches/Agents were notified of the action via email on December 23, 2015. Urgent Medical Device Recall Notification Letters/Urgent Medical Device Recall Notification Acknowledgement Forms dated December 23, 2015 were sent to Branches/Agents via UPS on December 28, 2015. Urgent Medical Device Recall Notification Letters/Urgent Medical Device Recall Notification Acknowledgement Forms dated December 23, 2015 were sent to Hospital Risk Managers and doctors via UPS on 12/23/2015. The notification instructed customers of the related issue with the affected product; how to identify affected product; the potential hazards associated with affected product; risk mitigations; and actions needed to be taken. Customers were asked to complete and return the attached Product Recall Acknowledgement Form within 5 days and either email ([email protected] or Fax (1-866-672-0627 ) the response form back. Customers were instructed to return all affected products to Stryker C/O Stericycle, 2670 Executive Drive, Suite A, Indianapolis, IN 46241 (Attn RA2015-120-Event 7808). A point of contact was provided in case the customer had any questions 201.831.5272 .
- Quantity in commerce
- 3,741 units
- Distribution
- Nationwide Distribution.
- Product codes and lots
- All lots of the following Catalog Numbers: 2704-1-024; 2704-3-018; 6703-1-018; 6703-2-120; 6704-1-018; 6704-2-120 and 6704-3-120.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0855-2016 | Class II | Terminated | Voluntary: Firm initiated |