Z-0855-2016Class II

Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures

Howmedica Osteonics Corp / initiated 2015-12-23 / Terminated / root cause: labeling and UDI

Stryker Howmedica Osteonics Corp. began a recall of Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures on . FDA classified it as Class II and gives the reason as "The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be resterilized. Since the IFU...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0855-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Vitallium Wire (6 Pack) Non Sterile. For use in bone procedures.
Product code
JDR Staple, Fixation, Bone
Reason for recall
The wire packages are correctly marked with a "NON-STERILE" label, however, the enclosed Instructions For Use (IFU) states the device are sterilized via gamma irradiation and should not be resterilized. Since the IFU states the devices are sterile, no instructions for moist heat sterilization are provided.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Stryker Branches/Agents were notified of the action via email on December 23, 2015. Urgent Medical Device Recall Notification Letters/Urgent Medical Device Recall Notification Acknowledgement Forms dated December 23, 2015 were sent to Branches/Agents via UPS on December 28, 2015. Urgent Medical Device Recall Notification Letters/Urgent Medical Device Recall Notification Acknowledgement Forms dated December 23, 2015 were sent to Hospital Risk Managers and doctors via UPS on 12/23/2015. The notification instructed customers of the related issue with the affected product; how to identify affected product; the potential hazards associated with affected product; risk mitigations; and actions needed to be taken. Customers were asked to complete and return the attached Product Recall Acknowledgement Form within 5 days and either email ([email protected] or Fax (1-866-672-0627 ) the response form back. Customers were instructed to return all affected products to Stryker C/O Stericycle, 2670 Executive Drive, Suite A, Indianapolis, IN 46241 (Attn RA2015-120-Event 7808). A point of contact was provided in case the customer had any questions 201.831.5272 .
Quantity in commerce
3,741 units
Distribution
Nationwide Distribution.
Product codes and lots
All lots of the following Catalog Numbers: 2704-1-024; 2704-3-018; 6703-1-018; 6703-2-120; 6704-1-018; 6704-2-120 and 6704-3-120.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0855-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0855-2016Class IITerminatedVoluntary: Firm initiated