Olympus TJF-Q180V flexible gastrointestinal Duodenoscope
Olympus Corporation of the Americas began a recall of Olympus TJF-Q180V flexible gastrointestinal Duodenoscope Product Usage: This instrument has been designed to be used with an Olympus video system center, light... on . FDA classified it as Class II and gives the reason as "Olympus America Inc is conducting a voluntary removal/corrective action of all TJF-Q180V duodenoscopes in order to replace the forceps elevator mechanism Olympus is replacing the forceps elevator mechanism with a new...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0757-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus TJF-Q180V flexible gastrointestinal Duodenoscope Product Usage: This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
- Product code
- FDT Duodenoscope And Accessories, Flexible/Rigid
- Reason for recall
- Olympus America Inc is conducting a voluntary removal/corrective action of all TJF-Q180V duodenoscopes in order to replace the forceps elevator mechanism Olympus is replacing the forceps elevator mechanism with a new forceps elevator design consistent with the design specification in the recently cleared TJF-Q180V 510k.
- Root cause tag
- design
- FDA root cause
- No Marketing Application
- Action
- Olympus America Inc. sent an Urgent Notification letter dated January 15, 2016 to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory for the affected product, Olympus will contact facilities to make arrangements for return of the TJF-Q180V duodenoscope(s) for the elevator mechanism replacement. For questions call (484) 896-5688.
- Quantity in commerce
- 4436
- Distribution
- US Nationwide Distribution except PR
- Product codes and lots
- all codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0757-2016 | Class II | Terminated | Voluntary: Firm initiated |