BioMerieux SA began a recall of API ZYM Bx2 (Ref 70493) Product Usage: ZYM B reagent is an additional test used for revealing the results of some miniaturized biochemical test including in... on . FDA classified it as Class II and gives the reason as "Discrepant identification results related to API identification strips. These discrepant identification results are due to false negative results on the tests of the strips linked to the ZYM B reagent". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0756-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- API ZYM Bx2 (Ref 70493) Product Usage: ZYM B reagent is an additional test used for revealing the results of some miniaturized biochemical test including in the involved API strips.
- Product code
- LIB Device, General Purpose, Microbiology, Diagnostic
- Reason for recall
- Discrepant identification results related to API identification strips. These discrepant identification results are due to false negative results on the tests of the strips linked to the ZYM B reagent.
- FDA root cause
- Other
- Action
- BioMrieux sent an Urgent Field Safety Notice letter DATED jANUARY 8, 2015 to affected customers. The letter identified the affected product, problem and action to be taken. dated . The letter indicated that they could continue to use product in inventory with mandatory requirements that before using a ZYM B ampule and each day of use of API strips that they perform a quality control with ATCC strains as described in each package insert of involved API strips. The letter also requested that they distribute this letter to all appropriate personnel in the laboratory, retain a copy for their files, and forward the information to all parties that may use the product. For additional assistance or have any questions, contact your local bioMrieux Customer Service representative.
- Quantity in commerce
- 13,271
- Distribution
- Nationwide Distribution in the states of: AL, CA, CO, CT, DC, FL, GA, HI, IL, IN, IA, KS, KY, ME, MD, MA, MI, MN, MS, MO, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TX, VT, VA, WA, WV, WI, WY
- Product codes and lots
- Lot No. PF - 1003765610/Lot No. Ampules - 1003693040/Exp. date 1/15/16 Lot No. PF - 1003772410/Lot No. Ampules - 1003711170/Exp. date 1/19/16 Lot No. PF - 1003772470/Lot No. Ampules - 1003733320/Exp. date 1/29/16 Lot No. PF - 1003787970/Lot No. Ampules - 1003693020/Exp. date 1/12/16 Lot No. PF - 1003790030/Lot No. Ampules - 1003790010/Exp. date 2/26/16 Lot No. PF - 1004077210/Lot No. Ampules - 1004024590/Exp. date 5/25/16 Lot No. PF - 1004097610/Lot No. Ampules - 1004040410/Exp. date 6/2/16 Lot No. PF - 1004355410/Lot No. Ampules - 1004306800/Exp. date 9/20/16 Lot No. PF - 1004355730/Lot No. Ampules - 1004306790/Exp. date 9/20/16
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0756-2016 | Class II | Terminated | Voluntary: Firm initiated |