Z-1240-2016Class II
Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757
Siemens Healthcare Diagnostics Inc. / initiated 2016-01-13 / Terminated / root cause: design
Siemens Healthcare Diagnostics, Inc began a recall of Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the... on . FDA classified it as Class II and gives the reason as "The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1240-2016
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens ADVIA Centaur Enhance Estradiol Test Code eE2 Catalog Number: 10490889, 10491445, 10697757 Product Usage: For in vitro diagnostic use in the quantitative determination of estradiol in human serum and plasma (heparinized and EDTA) using the ADVIA Centaur, ADVIA Centaur XP, ADVIA Centaur XPT and ADVIA Centaur CP systems.
- Product code
- CHP Radioimmunoassay, Estradiol
- Reason for recall
- The drug fulvestrant (Faslodex) may cause falsely elevated estradiol results.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC) to affected Siemens Healthcare Diagnostics customers in the United States on January 13, 2016 to be delivered to customers on January 14, 2016 via Federal Express. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 13, 2016. The letters identified affected product, problem and the actions to be taken. Customers may continue to use the estradiol assays to report patient results however, the assays should not be used for patients being treated with fulvestrant. Customers where instructed to complete and return the Correction Effectiveness Check attached. For questions contact your Siemens Customer Care Center or your local Siemens technical support representative.
- Quantity in commerce
- 33604 units
- Distribution
- Worldwide Distribution US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1240-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1241-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1242-2016 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1243-2016 | Class II | Terminated | Voluntary: Firm initiated |